Relative Humidity Range: 10 to 95% (limited by a 7°C dew point)
Relative Humidity Constancy: ± 0.5%
Relative Humidity Uniformity: ± 1.0% at 25° / 60%
Temperature Range: 10 to 60°C
Temperature Constancy: ± 0.2%
Temperature Uniformity: ± 0.2% at 25° / 60%
The Parameter 40 cubic foot chamber features a unique humidity control system that enables the user to attain unsurpassed levels of RH/temperature control and uniformity. The chamber has been designed with an exceptional heat exchange system that takes the refrigeration coil out of the process air stream to enhance the long-term performance of this chamber.
These test chambers are designed to accurately produce temperature and humidity conditions required for applications including stability studies, package testing, TAPPI, MIL-SPEC, and vapor transmission.